Current¶
SCPI Commands
FETCh:CDMA:MEASurement<Instance>:MEValuation:LIST:ACP:EXTended:CURRent
CALCulate:CDMA:MEASurement<Instance>:MEValuation:LIST:ACP:EXTended:CURRent
- class Current[source]
Current commands group definition. 2 total commands, 0 Sub-groups, 2 group commands
- class CalculateStruct[source]
Response structure. Fields:
Reliability: int: decimal ‘Reliability Indicator’. In list mode, a zero reliability indicator indicates that the results in all measured segments are valid. A non-zero value indicates that an error occurred in at least one of the measured segments.
Seg_Reliability: List[int]: 0 | 3 | 4 | 8 The segment reliability indicates whether one of the following exceptions occurred in this segment: 0: No error 3: Signal overflow 4: Signal low 8: Synchronization error If a combination of exceptions occurs, the most severe error is indicated.
Acp: List[float]: No parameter help available
Ms_Power_Wide: List[float]: float Results of the wideband (8 MHz) power measurement. Range: -55 dBm to 42 dBm , Unit: dBm
Ms_Power_Narrow: List[float]: float Results of the narrowband (1.23 MHz) power measurement. Range: -55 dBm to 42 dBm , Unit: dBm
Out_Of_Tol_Count: List[float]: decimal Out of tolerance result, i.e. percentage of measurement intervals of the statistic count exceeding the specified limits. Range: 0 % to 100 % , Unit: %
Cur_Stat_Count: List[float]: decimal Number of evaluated valid slots in this segment. Range: 0 to 1000
- class FetchStruct[source]
Response structure. Fields:
Reliability: int: decimal ‘Reliability Indicator’. In list mode, a zero reliability indicator indicates that the results in all measured segments are valid. A non-zero value indicates that an error occurred in at least one of the measured segments.
Seg_Reliability: List[int]: 0 | 3 | 4 | 8 The segment reliability indicates whether one of the following exceptions occurred in this segment: 0: No error 3: Signal overflow 4: Signal low 8: Synchronization error If a combination of exceptions occurs, the most severe error is indicated.
Acp: List[float]: No parameter help available
Ms_Power_Wide: List[float]: float Results of the wideband (8 MHz) power measurement. Range: -55 dBm to 42 dBm , Unit: dBm
Ms_Power_Narrow: List[float]: float Results of the narrowband (1.23 MHz) power measurement. Range: -55 dBm to 42 dBm , Unit: dBm
Out_Of_Tol_Count: List[float]: decimal Out of tolerance result, i.e. percentage of measurement intervals of the statistic count exceeding the specified limits. Range: 0 % to 100 % , Unit: %
Cur_Stat_Count: List[int]: decimal Number of evaluated valid slots in this segment. Range: 0 to 1000
- calculate() → CalculateStruct[source]
# SCPI: CALCulate:CDMA:MEASurement<Instance>:MEValuation:LIST:ACP:EXTended:CURRent value: CalculateStruct = driver.multiEval.listPy.acp.extended.current.calculate()
Returns the adjacent channel power (ACP) results of each active segment (see method RsCmwCdma2kMeas.Configure.MultiEval. ListPy.value) . If enabled the wideband power and narrowband power results are returned, too. To define the statistical length for AVERage, SDEviation, MINimum and MAXimum calculation and to enable the calculation of the results, use the command method RsCmwCdma2kMeas.Configure.MultiEval.ListPy.Segment.Spectrum.set. The values described below are returned by FETCh commands. CALCulate commands return limit check results instead, one value for each result listed below. The values listed below in curly brackets {} are returned for each active segment: {…}seg 1, {…}seg 2, …, {…}seg n. The number of active segments n is determined by method RsCmwCdma2kMeas.Configure.MultiEval.ListPy.count. The number to the left of each result parameter is provided for easy identification of the parameter position within the result array.
- return
structure: for return value, see the help for CalculateStruct structure arguments.
- fetch() → FetchStruct[source]
# SCPI: FETCh:CDMA:MEASurement<Instance>:MEValuation:LIST:ACP:EXTended:CURRent value: FetchStruct = driver.multiEval.listPy.acp.extended.current.fetch()
Returns the adjacent channel power (ACP) results of each active segment (see method RsCmwCdma2kMeas.Configure.MultiEval. ListPy.value) . If enabled the wideband power and narrowband power results are returned, too. To define the statistical length for AVERage, SDEviation, MINimum and MAXimum calculation and to enable the calculation of the results, use the command method RsCmwCdma2kMeas.Configure.MultiEval.ListPy.Segment.Spectrum.set. The values described below are returned by FETCh commands. CALCulate commands return limit check results instead, one value for each result listed below. The values listed below in curly brackets {} are returned for each active segment: {…}seg 1, {…}seg 2, …, {…}seg n. The number of active segments n is determined by method RsCmwCdma2kMeas.Configure.MultiEval.ListPy.count. The number to the left of each result parameter is provided for easy identification of the parameter position within the result array.
- return
structure: for return value, see the help for FetchStruct structure arguments.