Oltr

SCPI Commands

CONFigure:CDMA:MEASurement<Instance>:OLTR:TOUT
CONFigure:CDMA:MEASurement<Instance>:OLTR:REPetition
CONFigure:CDMA:MEASurement<Instance>:OLTR:SEQuence
CONFigure:CDMA:MEASurement<Instance>:OLTR:MOEXception
class Oltr[source]

Oltr commands group definition. 11 total commands, 4 Sub-groups, 4 group commands

get_mo_exception()bool[source]
# SCPI: CONFigure:CDMA:MEASurement<Instance>:OLTR:MOEXception
value: bool = driver.configure.oltr.get_mo_exception()

Specifies whether measurement results that the R&S CMW identifies as faulty or inaccurate are rejected.

return

meas_on_exception: OFF | ON ON: Results are never rejected OFF: Faulty results are rejected

get_repetition()RsCmwCdma2kMeas.enums.Repeat[source]
# SCPI: CONFigure:CDMA:MEASurement<Instance>:OLTR:REPetition
value: enums.Repeat = driver.configure.oltr.get_repetition()

Specifies the repetition mode of the measurement. The repetition mode specifies whether the measurement is stopped after a single shot or repeated continuously. Use CONFigure:..:MEAS<i>:…:SCOunt to determine the number of measurement intervals per single shot.

return

repetition: SINGleshot | CONTinuous SINGleshot: Single-shot measurement CONTinuous: Continuous measurement

get_sequence()int[source]
# SCPI: CONFigure:CDMA:MEASurement<Instance>:OLTR:SEQuence
value: int = driver.configure.oltr.get_sequence()

Sets/gets the number of measurement sequences within a single OLTR measurement. Each sequence consists of a power UP or power DOWN step, followed by a power step in the opposite direction (see method RsCmwCdma2kMeas.Configure.Oltr.Pstep. direction.

return

no_of_meas_seq: numeric Range: 1 to 5

get_timeout()float[source]
# SCPI: CONFigure:CDMA:MEASurement<Instance>:OLTR:TOUT
value: float = driver.configure.oltr.get_timeout()

Defines a timeout for the measurement. The timer is started when the measurement is initiated via a READ or INIT command. It is not started if the measurement is initiated manually ([ON | OFF] key or [RESTART | STOP] key) . When the measurement has completed the first measurement cycle (first single shot) , the statistical depth is reached and the timer is reset. If the first measurement cycle has not been completed when the timer expires, the measurement is stopped. The measurement state changes to RDY. The reliability indicator is set to 1, indicating that a measurement timeout occurred. Still running READ, FETCh or CALCulate commands are completed, returning the available results. At least for some results, there are no values at all or the statistical depth has not been reached. A timeout of 0 s corresponds to an infinite measurement timeout.

return

timeout: numeric Unit: s

set_mo_exception(meas_on_exception: bool)None[source]
# SCPI: CONFigure:CDMA:MEASurement<Instance>:OLTR:MOEXception
driver.configure.oltr.set_mo_exception(meas_on_exception = False)

Specifies whether measurement results that the R&S CMW identifies as faulty or inaccurate are rejected.

param meas_on_exception

OFF | ON ON: Results are never rejected OFF: Faulty results are rejected

set_sequence(no_of_meas_seq: int)None[source]
# SCPI: CONFigure:CDMA:MEASurement<Instance>:OLTR:SEQuence
driver.configure.oltr.set_sequence(no_of_meas_seq = 1)

Sets/gets the number of measurement sequences within a single OLTR measurement. Each sequence consists of a power UP or power DOWN step, followed by a power step in the opposite direction (see method RsCmwCdma2kMeas.Configure.Oltr.Pstep. direction.

param no_of_meas_seq

numeric Range: 1 to 5

set_timeout(timeout: float)None[source]
# SCPI: CONFigure:CDMA:MEASurement<Instance>:OLTR:TOUT
driver.configure.oltr.set_timeout(timeout = 1.0)

Defines a timeout for the measurement. The timer is started when the measurement is initiated via a READ or INIT command. It is not started if the measurement is initiated manually ([ON | OFF] key or [RESTART | STOP] key) . When the measurement has completed the first measurement cycle (first single shot) , the statistical depth is reached and the timer is reset. If the first measurement cycle has not been completed when the timer expires, the measurement is stopped. The measurement state changes to RDY. The reliability indicator is set to 1, indicating that a measurement timeout occurred. Still running READ, FETCh or CALCulate commands are completed, returning the available results. At least for some results, there are no values at all or the statistical depth has not been reached. A timeout of 0 s corresponds to an infinite measurement timeout.

param timeout

numeric Unit: s

Cloning the Group

# Create a clone of the original group, that exists independently
group2 = driver.configure.oltr.clone()